Horizontal Adjustment:
Horizontal adjustment is performed as illustrated in Fig. 5.1 and 5.2.
Gently loosen the screw xing the chamber and carefully move the myograph
chamber in either clockwise or counter clockwise direction to adjust the horizontal
alignment of the glass cannulas.
Horizontal Adjustment Screw.
Loosen this screw and carefully turn CW or CCW. Observe
how the pippettes move in horisontal plane.
Figure 5.2 Horizontal adjustment of glass pipettes
Vertical Adjustment:
Vertical adjustment is performed as illustrated in Fig. 5.1.
Gently turn the black dish, underneath the left glass cannula, CW or CCW to adjust
the vertical alignment of the glass cannulas.
Longitudinal Adjustment:
Longitudinal adjustment of the glass cannulas is performed using the micro
positioner on the right side of the culture myograph unit.
42
Culture Myograph System Model 202CM